AC-POWER
Issue 1 June 2005 337
1 (cont’d) FAIL
Procedures for a J58890CE
1. Check for AC power at the wall outlet (99–127 or 200–250 VAC).
2. If AC power is absent, then the problem is not with the system itself.
Power must be restored to the wall outlet.
If present, there could be a problem with the battery or battery
chargers. Resolve every alarm logged against POWER (battery &
battery charger), and rerun the test.
3. If the test still fails, then fuse F5 may have opened. Replace this
fuse, and rerun the test.
!
WARNING:
WARNING: Turn off the 397C Battery Charger before inserting or
removing Fuse F5.
4. If the test still fails, then the server’s watchdog or a PN’s
Maintenance circuit pack may be incorrectly reporting the problem.
Resolve every alarm against these MOs, and rerun the test. Some
watchdog and Maintenance circuit-pack errors may not be detected
by their respective maintenance routines, but can cause many
environment tests to fail.
5. If more than two environment tests are failing, replace the suspected
circuit pack (depending on the system configuration), and rerun the
test.
100 FAIL The cabinet is currently without AC power, and AC power maintenance
is in the middle of timing the NPH interval.
1. Follow the repair steps for error code 1.
PASS The server’s watchdog or the PN’s Maintenance circuit pack has reported
that the cabinet has AC power.
Table 88: Test #78 Power Query Test (continued)
Error
Code
Test
Result
Description / Recommendation
2 of 2