XXX-BD (Common Port Circuit Pack/Media Module)
Issue 1 June 2005 2549
Ringing Application Test (#51)
This test checks the ringing application circuitry common to every port on an Analog Line circuit
pack.
Table 910: Test #51 Ringing Application Circuit Test
Error
Code
Test
Result
Description / Recommendation
ABRT Internal system error. Could not allocate the necessary system resources
to run test. The circuit pack is not installed.
1000
2100
ABRT Could not allocate the necessary system resources to run test.
1. Retry the command at 1-minute intervals up to 5 times.
1004 ABRT The port was seized by a valid call during the test. The test has been
aborted.
1. Use display port location to determine the station extension.
Use status station to determine the service state of the port. If
the service state indicates that the port is in use, then the port is
unavailable for certain tests. (Refer to Chapter 8, for a full description
of all possible states). Wait until the port is idle before retesting.
2. Retry the command at 1-minute intervals up to 5 times.
1008 ABRT Could not allocate a ringing circuit for one of the following reasons: all the
ringing circuits are in use; the ringing generator is defective; ringing
generator is not wired correctly.
1. If the test continues to abort, look for RING-GEN error in Error Log.
a. If there are RING-GEN errors, see RING-GEN (Analog Ring
Generator) on page 1942 and try to resolve any problem(s). Then,
go to step 2.
b. If there are no RING-GEN errors, and the test continues to abort,
enter test board location on other TN742, TN769, or TN746
Analog circuit packs. If an ABRT with Error Code 1008 occurs for
this test on other circuit packs as well, then the ringing generator
may be defective or may not be wired properly. See RING-GEN
(Analog Ring Generator) on page 1942 maintenance
documentation for details. If an ABRT with Error Code 1008 does
NOT occur on the other ports, then all four ring generators are in
use. Exit from this documentation.
2. Retry the command.
1 of 2