Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG2-20
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
Fig. 2-13: X-STREAM X2FD - Power Supply Terminals / Fuse Holders
3
2
4
1
Power supply
Nominal input voltage 100–240 V 50/60 Hz,
global power supply unit
Input voltage range 85–264 V , 47–63 Hz
Nominal input current
standard 1.3–0.7 A max.
with temperature control 3–1.5 A max.
Connection via touch proof screw-type
terminals located inside the unit near
cable glands (Fig. 2-13).
Cable cross-section: max. 4 mm
2
(10 AWG), end sleeves not
required.
Cable entry via 1 cable gland, classied IP 68 or suitable
conduit with metric-to-NPT adaptor
Permissible cable outer diameter: 3–13 mm (0.11–0.5 inch),
dependent on inset used in cable gland
Power supply fuses
Fuse holders are integrated within power
supply terminals.
Fuse ratings: AC 230 V / T 4 A / 5x20 mm
2.2.4 Model-Specic Technical Data: X-STREAM X2FD
1 Power terminals with integrated
fuse holders
2 Protective earth terminal (PE)
3 Power cable entry
4 EMI power supply lter