7
Maintenance & Procedures
Seal the open analyzer´s gas t-
tings utilizing PVC caps to avoid
pollution of inner gas path.
Moisten the lint-free cloth with a mixture of
3 parts of water and 1 part of the general
purpose detergent.
7.6 Cleaning the Instrument´s Outside
7.6 Cleaning the Instrument´s Outside
Use a liquid general purpose detergent and
a lint-free cloth for cleaning the analyzer´s
outside.
Procedure
Disconnect the instrument from power!
If disconnecting from gas lines is re-
quired, take care of the following:
EXPLOSIVE, FLAMMABLE AND HARMFUL GASES HAZARD
Before opening gas paths they must be purged with ambient air or neutral
gas (N
2
) to avoid hazards caused by toxic, ammable, explosive or harmful
to health sample gas components!
HAZARD FROM UNHEALTHY SUBSTANCES
Take care to follow the safety instructions and instructions for use given
by the manufacturer of the chosen general purpose detergent!
Do NOT drench the cloth, just
moisten it to prevent liquid en-
tering the housing!
Clean the analyzer housing outside with
the moistened cloth.
If need be dry the housing after cleaning.