Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 5-19
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
5
Startup
The rst line allows the user to enter the cur-
rent air pressure manually when no pressure
sensor is installed, or to read the current
pressure when a sensor is installed (
INSTALLED OPTIONS menu). The measu-
rement unit is set in the DISPLAY SETUP
menu.
If no pressure sensor is installed, enter the
current air pressure here and adjust it when
signicant changes take place: this improves
the accuracy of the instrument.
Signal damping (set in the DAMPING menu)
allows the smoothing of the measuring signal,
but also affects the reactio time of outputs and
display. The factory setting is 0 seconds. and
any value between 0 and 28 seconds can be
set. In multi-channel units, the value for each
channel must be entered separately.
Pressure: 1014.0 hPa
Damping..
Press the LEFT key to return to the SETUP
menu and from there open the MEASURE-
MENT MENU.
Setup..
Measurement..
5.7 Checking the Settings