Emerson Process Management GmbH & Co. OHG 6-29
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6
Software Menus
Damping: 5 s
6.2.3.3.1 Signal Damping Settings
The time for the electronic signal damping is
entered here.
Lower values give results which are updated
more quickly, higher values supress noise
from varying gas concentrations.
Possible values: 0 .. 28 seconds
6.2 Menu System - Measurement Setup
Multi-channel unit:
Select the component to be set in the SELECT
COMPONENT menu.
Setup..
Measurement..
Damping..
Note!
The analyzer’s total delay time (t
90
time) is
the sum of the signal damping time and the
physical time lag caused e.g. by the properties
of the gas ow and the sensors.
Multi-channel unit:
Press the key to enter the SELECT COM-
PONENT menu to change the settings for a
different channel..