Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 6-5
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6
Software Menus
6.2 Menu System - Control Menu
6.2.2 Control Menu
Control..
If the system is set up accordingly, the access
code for level 1 must be entered to gain ac-
cess to this menu.
Lock menus!
Acknowledgements..
Pump: Off
Zoom..
Available when internal pump installed:
Switch internal pump On or Off.
Note!
This line does not appear when a digital input
is used to control the pump (
6.2.3.4.4,
page 6-39) or when no internal pump is
available.
Pressing the
8 key in this line will lock all
menus for which the code parameter in the
menu access settings has been set to On or
1 Min ( 6.2.3.1.2, page 6-18)
Page 2
Zoom submenu
6.2.2.6, page 6-13
Note!
The zoom function for at least 1 analog output
must be activated for this line to appear (
6.2.3.4.1, page 6-31).
Acknowledgements submenu
6.2.2.5, page 6-12
Zero calibration menu
6.2.2.1, page 6-6
Span calibration menu
6.2.2.2, page 6-7
Apply gas menu
6.2.2.4, page 6-11
Advanced calibration menu
6.2.2.3, page 6-10
Zero calibration..
Span calibration..
Adv. calibration..
Apply gas..
Page 1
Note!
Internal or internal valves must be installed for
the last two lines to appear (
6.2.3.4.4,
page 6-39)