9.3 List of Parameters and Registers - Sorted by Tag Name
Tag Name
Address
Data
Type
Client
Access
Description
Modicon Daniel
Service.AnalogOut.AOutAdjustStart4 3308 3308 Short R/W
offset value for ne adjustment of ana-
log output4
Service.Communication.SvcModbusFt32 3028 3028 Word R/W
Svc: 32bit regs format (0=Daniel,
1=Modicon, 2=Modicon swap)
Service.General.ChannelId1
3201…
3210
3201…
3210
String R/W channel identication text of comp1
Service.General.ChannelId2
3211…
3220
3211…
3220
String R/W channel identication text of comp2
Service.General.ChannelId3
3221…
3230
3221…
3230
String R/W channel identication text of comp3
Service.General.ChannelId4
3231…
3240
3231…
3240
String R/W channel identication text of comp4
Service.General.EmersonAccCode
3186…
3189
3186…
3189
String R/W
code for getting LOI access to Emerson
areas
Service.General.EmersonAccMode 3190 3190 Word RO
mode for getting LOI access to Emer-
son areas
Service.General.NumberChannels 3001 3001 Word R/W number of built-in component channels
Service.General.SerialNumber
3196…
3200
3196…
3200
String R/W serial number of the device
Service.General.WarmupTime 3002 3002 Word R/W
time in secs that is used for warmup
phase
Service.InstalledOptions.AOutNumber 3024 3024 Word R/W number of installed analog outputs
Service.InstalledOptions.DIO_Installed 3030 3030 Word R/W
dig. IO installed: 0=None, 1=Card1,
2=Card1+2
Service.InstalledOptions.FlowAlarmInstalled 3023 3023 Word R/W digital ow alarm installed
Service.InstalledOptions.FlowSensorAssi-
gned1
3273 3273 Word R/W
ow sensor assigned to comp1
(0=None, etc.)
Service.InstalledOptions.FlowSensorAssi-
gned2
3274 3274 Word R/W
ow sensor assigned to comp2
(0=None, etc.)
Service.InstalledOptions.FlowSensorAssi-
gned3
3275 3275 Word R/W
ow sensor assigned to comp3
(0=None, etc.)
Service.InstalledOptions.FlowSensorAssi-
gned4
3276 3276 Word R/W
ow sensor assigned to comp4
(0=None, etc.)
Service.InstalledOptions.PressureSensorIn-
stalled
3027 3027 Word R/W
pressure (0=manual,1=intSens,2=cycl
Remote)
Service.InstalledOptions.PumpInstalled 3043 3043 Word R/W
pump is installed and controlled by
device
Service.InstalledOptions.SIntInstalled 3025 3025 Word R/W serial interface hardware installed
Service.InstalledOptions.TempSensorAssi-
gned1
3277 3277 Word R/W
temp sensor assigned to comp1
(0=None, etc.)
Service.InstalledOptions.TempSensorAssi-
gned2
3278 3278 Word R/W
temp sensor assigned to comp2
(0=None, etc.)