9.2 Supported Functions
Modbus Function
Function Code
Note
1)
decimal (hex)
ReadCoils 01 (0x01) for registers of 2000
ReadDiscreteInputs 02 (0x02) for registers of 1000
ReadHoldingRegisters 03 (0x03) for registers of 3000, 8000, 9000
ReadInputRegisters 04 (0x04) for registers of 4000, 8000, 9000
WriteSingleCoil 05 (0x05) for registers of 2000
WriteSingleRegister 06 (0x06) for registers of 3000
Diagnostic 08 (0x08) sub function "00 = Return Query Data" only
WriteMultipleCoils 15 (0x0F) for registers of 2000
WriteMultipleRegisters 16 (0x10) for registers of 3000, 8000, 9000
EncapsulatedInterfaceTransport 43 (0x28)
sub function "0x60" and "0x81" only
(to be used for conguration le transfer)
9.2 Modbus - Supported Functions
1)
Registers ranges 8000 and 9000 are Daniel long word or oating point registers.
To calculate the related Modicon registers use the following table:
Daniel Modicon Data type
8001 - 8499 equals 5001 - 5999 long word
9001 - 9999 equals 6001 - 7999 oating point
or the following pages for comparisons of all Daniel and Modicon registers.
9.3 List Of Parameters And Registers - Sorted By Tag Name
Note!
The client access column in the following list provides information about the read only (RO) or
read/write (R/W) access of each parameter.
All parameters with read/write access and tag names beginning with "Service." require entering
the service level access code into the register 3008 for parameter "Service.RemoteSecurity" to
enable write access.