Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG5-32
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
• Dening high pre-alarm and main alarm
Standard mode:
An alarm results in the assigned relay being
activated.
Settings:
• Level 1 > Level 2
• Level 1-Function: High
• Level 2-Function: High
If an alarm is active, a corresponding mes-
sage is displayed in line 4 of the measurement
display.
Failsafe mode:
An alarm results in the assigned relay being
deactivated.
Settings:
• Level 1 > Level 2
• Level 1-Function: High FS
• Level 2-Function: High FS
Fig. 5-3: High Pre-Alarm and Main Alarm
5.7 Checking the Settings
If two upper limits are set with one limit hig-
her than the other (g. 5-3), a pre-alarm is
triggered when the measured concentration
exceeds the rst limit (area B). If no corrrecti-
ve measures are taken and the concentration
exceeds the second limit (area C), a main
alarm is triggered.
Up to two alarms may be active per channel
at any one time.
Concentration
Activated
Time
Concentration level
Pre-alarm limit
Main alarm limit
Deactivated
Activated
Deactivated
Activated
Deactivated
Activated
Deactivated
A B C
On
Off
Relay condition
Standard mode
Failsafe mode
Status message on front panel
Main alarm rising
Pre-alarm rising
Main alarm rising
Pre-alarm rising