Technical data
Interfaces, signal inputs and outputs
Interface signals are accessed in different ways depending on the analyzer model. (This does
not apply to Ethernet, which always uses RJ45):
X-STREAM X2GK, X2GP: standard: subminD plugs and sockets
optional: screw-type terminal adapters
X-STREAM X2XF, X2FD: internal screw-type terminals
All models are supplied with
up to 4 analog outputs 4 (0)–20 mA (R
B
≤ 500 Ω)
(Standard: 1 analog output per channel) electrically isolated from each other and
from other electronic components;
user-congurable activation and
deactivation concentration levels;
support for NAMUR NE 43 operation
modes
, congurable via keypad and
Modbus
4 relay outputs dry contacts,
max. load. 30 V; 1 A; 30 W resistive
Each output can be congured to one of the
following functions: NAMUR NE 107 status signal
“Failure”
“Maintenance request”
“Out of specication”
“Function check”
1 of 2 concentration limits per channel,
Control signals for
external valve V1…V8,
external sample gas valve
external pump
Zoom status for analog outputs
2.1 Common Technical Data