7.7.1 Save CfgData to UserData
2
nd
menu page
FactData > CfgData..
CfgData > UserData..
UserData > CfgData..
Save-Load
CfgData > SvcPort!
SvcPort > CfgData..
Verify!
1
st
menu page
CfgData>UserData
Are you sure?
No!
Yes!
(i)
-COMMAND EXECUTED-
Setup..
Save-Load..
Copying data
- PLEASE WAIT -
Procedure X:E000
Starting at the measurement screen press the
DOWN key to open the MAIN MENU, enter
the SETUP and next the SAVE-LOAD.. menu.
Now select the "CfgData>UserData.." line and
press the ENTER key.
Press the DOWN key to open the second
menu page.
A screen appears to conrm the operation:
Select Yes! and after pressing the ENTER
key a new screen comes up showing the
current status.
The instrument now stores the currently used
(and changed by operator) analyzer setup
into a special memory area. This data is
then called UserData and used for backup
only, while the data used for operation still is
CfgData.
Any further changes affecting the instrument's
setup update the CfgData only, as long as not
manually saved to the UserData set.
When the procedure has nished, the COM-
MAND EXECUTED screen shows up.
If system is setup accordingly, access level 3
code must be entered to gain access to this
menu.
7.7 Save / Restore Conguration Data Sets