Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 6-7
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6
Software Menus
Cancel calibration!
Start calibration!
Span gas 2000.0 ppm
CO2.1 134.1 ppm
6.2.2.2 Span Calibration Menu
6.2 Menu System - Control Menu
Note!
For further details on calibration procedures,
see
chapter 7 Maintenance.
Control..
Span calibration..
Multi-channel unit:
Select the channel to be calibrated in the
SELECT COMPONENT menu.
Pressing the 8 key in this line will abort the
span calibration and cancel any changes.
Pressing the
8 key in this line will start the
span calibration.
Current nominal span gas concentration
(span gas concentration can be set in the
SETUP menu
6.2.3.2.1, page 6-24)
Current measured gas concentration.
Page 1
Page 2
RESET..
Status..
CO2.1 134.1 ppm
Current measured gas concentration.
Press the 8 key in this line to view the current
calibration status ( 6.2.2.2.2, page 6-9)
Multi-channel unit:
Press the key to enter the SELECT COM-
PONENT menu to change the settings for a
different channel.
Press the
8 key in this line to enter a subme-
nu to reset the calibration parameters (
6.2.2.2.1, page 6-8)