Interval time
ZeroAll: 2 h
ZSCalAll: 10 h
AutoCal in..
6.2 Menu System - Calibration Setup
6.2.3.2.3 Interval Time Menu
If the system is set up accordingly, the access
code for level 3 must be entered to gain ac-
cess to this menu.
Sets the time between two zero calibrations
in autocal mode.
Sets the time between two combined zero &
span calibrations in autocal mode.
Possible values for both parameters:
0 .. 999 h
Note!
This line appears if at least one interval time
has been set.
Specify a time interval to elapse from the
current time until the rst autocal starts
6.2.3.2.3.1, page 6-27
Note!
The following lines are only available when
the “Valves” parameter in the INSTALLED
OPTIONS menu is set to a value other than
none.
Setup..
Calibration..
Interval time..
Note!
For further details on calibration procedures,
see
chapter 7 Maintenance.