Emerson Process Management GmbH & Co. OHG6-32
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6.2 Menu System - Analog Outputs Menu
Page 2
Page 3
Available analog outputs (min. 1, max. 4) can
be set in greater detail via these submenus.
6.2.3.4.1.2, page 6-34
The lines on this page enable to congure the
behaviour of the analog outputs in case the
measured values are beyond the range limits:
Options:
Standard: The outputs are set to the values
marked * in table 6-1 below
Cong: The outputs can be adjusted to the
values marked ** in table 6-1 below.
Output1..
Output2..
Output3..
Output4..
Cut mode: Standard
Low cut: 3.80
High cut: 20.50
Output signal, if
"Signal-
Range“
Operation Mode
Failure
Signal Level
acc. NE 43
Measured
value is
valid
Measured value is
below lower range
limit
Measured value is
above upper range
limit
An internal
failure
occured
Cable is
broken
0-20 mA Dead-Zero - 0 ... 20 mA < -19 mA > 21.7 mA undened 0 mA
4-20 mA Live-Zero - 4 ... 20 mA < -19 mA > 21.7 mA undened 0 mA
0-20 mAL
similar Dead-
Zero
below 0 ... 20 mA
-0.2 mA*
(-1.8 … -0.01 mA)**
20.5 mA*
(20.01 ... 21.5 mA)**
-2 mA 0 mA
4-20 mAL similar Live-Zero below 4 ... 20 mA
3.8 mA*
(2.2 ...3.9 mA)**
20.5 mA*
(20.01 ... 21.5 mA)**
2 mA 0 mA
0-20 mAH
similar Dead-
Zero
above 0 ... 20 mA
-0.2 mA*
(-1.8 … -0.01 mA)**
20.5 mA*
(20.01 ... 21.5 mA)**
> 21.7 mA 0 mA
4-20 mAH similar Live-Zero above 4 ... 20 mA
3.8 mA*
(2.2 ...3.9 mA)**
20.5 mA*
(20.01 ... 21.5 mA)**
> 21.7 mA 0 mA
These two lines are editable only, if "Cut
mode" is set to Cong:
Low Cut: Congure the "below lower range
limit" output signal
High Cut: Congure the "above upper range
limit" output signal
Accepted values depend on the setting of
"SignalRange" and are marked ** in table 6-1.
Note!
These settings apply to ALL analog outputs!
Tab. 6-1: Analog Output Signals - Settings and Operational Modes