X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
5.7 Checking the Settings
Level1: 100 ppm
Function: Low
Level2: 500 ppm
Function: High
Press the LEFT key until the SETUP menu
is displayed, then select “Alarms” and open
the submenu. If you are using a multi-channel
analyzer, select the channel to be modied.
5.7.5 Setting Concentration Alarms
Two concentration limits can be set for each
channel. Valid settings for limit levels depend
on the measuring range and the value of the
“SpanRange” parameter (
next page):
An error message is displayed if an invalid
setting is input.
Note!
If concentration alarms are not being used,go
straight to page 5-34.
Should the measured concentration go bey-
ond one of the limits, a message is displayed
in the fourth line of the measurement display
and the corresponding digital output is acti-
vated if programmed to do so.