6.2 Menu System - Digital Outputs Menu
The following functions can be assigned to
each of the digital outputs (max. 22).
Label Description
Off Digital output unused
On Digital output constantly on
Test Switches the output on and off every second
V1 Switches external valve V1 on autocalibration
V2 Switches external valve V2 on autocalibration
V3 Switches external valve V3 on autocalibration
V4 Switches external valve V4 on autocalibration
V5 Switches external valve V5 on autocalibration
V6 Switches external valve V6 on autocalibration
V7 Switches external valve V7 on autocalibration
V8 Switches external valve V8 on autocalibration
Sample Switches external sample gas valve on autocalibration
Pump Switches an external pump
Failure NAMUR NE 107 “Failure” signal
MaintReq NAMUR NE 107 “Maintenance request” signal
OffSpec NAMUR NE 107 “Out of specication” signal
FctCheck NAMUR NE 107 “Function check” signal
Lim1Cmp1 Concentration alarm 1, component (channel) 1
Lim2Cmp1 Concentration alarm 2, component (channel) 1
Lim1Cmp2 Concentration alarm 1, component (channel) 2
Lim2Cmp2 Concentration alarm 2, component (channel) 2
Lim1Cmp3 Concentration alarm 1, component (channel) 3
Lim2Cmp3 Concentration alarm 2, component (channel) 3
Lim1Cmp4 Concentration alarm 1, component (channel) 4
Lim2Cmp4 Concentration alarm 2, component (channel) 4
Zoom1 Component (channel) 1 analog signal is zoomed
Zoom2 Component (channel) 2 analog signal is zoomed
Zoom3 Component (channel) 3 analog signal is zoomed
Zoom4 Component (channel) 4 analog signal is zoomed
FlowAlm Alarm by ow sensor
FLimAlm1 Alarm by ow monitor 1
FLimAlm2 Alarm by ow monitor 2
Functions can also be assigned to several
outputs simultaneously.
Tab. 6-2: Options for Digital Outputs