Technical data
Housing
Permissible operational ambient temperature: 0 °C to +50 °C (32 °F to 122 °F)
Permissible ambient temperature for storage: -20 °C to +70 °C (-4 °F to 158 °F)
Weight: approx. 12–16 kg (26.4–35.2 lb)
(dependent on conguration)
EN 60529 housing protection type: IP 20 for indoor use.
Units must be protected against dripping
water and splashes, and direct sunlight.
Gas connections: quantity: max. 8
optional 1 addit. purge gas connector
material: PVDF 6/4 mm
optionally stainless steel 6/4 mm or
1
4", other options on request
Power supply
Nominal input voltage 100–240 V 50/60 Hz,
global power supply unit
Input voltage range 85–264 V , 47–63 Hz
Nominal input current
standard 1.3–0.7 A max.
with temperature control 3–1.5 A max.
Connection via rubber connector
(IEC connector) with power switch at rear of unit ( gs. 2-5 & 2-6).
Power input fuses
The IEC connector incorporates holders for
two fuses.
Data for fuse links: AC 230 V / T 4 A / 5x20 mm