Startup
Signal: Comp-1
LowScale: 0 ppm
HighScale: 1000 ppm
Zoom..
The “Signal” parameter denes the value to
be sent to the selected output. The following
options (partly dependent on the number of
measuring channels and sensors installed)
are available:
Value Description
None
The analog signal is deactivated
0/4 mA
Either a 0 mA or 4 mA signal is generated, for
example to be used to test the processing in a
subsequent system. The actual type of generated
signal is setup in the previous menu in the “Si-
gnalRange” line ( previous page).
20 mA
A 20 mA signal is generated, with which, for exa-
mple, the processing of a signal can be tested.
Comp-1, Comp-2, Comp-3, Comp-4 Gas component of channel 1 to 4
Temp-1, Temp-2, Temp-3, Temp-4 Measured value from temperature sensor
Press-1, Press-2, Press-3, Press-4 Measured value from pressure sensor
Flow-1, Flow-2, Flow-3, Flow-4 Measured value from ow sensor
Zoom-C1, Zoom-C2, Zoom-C3, Zoom-C4
A “zoomed” signal is sent from the selected mea-
suring channel (C1 to C4). If one of these options
is selected, the “Zoom..” line appears in the menu
(see above), which allows a zoom to be set.
5.7 Checking the Settings