The number of gas connections and their
conguration may vary according to analyzer
version and installed options.
All gas connectors are labelled and can be
found on the
•
analyzer’s rear panel (X-STREAM X2GP,
X-STREAM X2GK)
• underside of the analyzer (X-STREAM eld
housings)
Should it be necessary to open the gas lines,
the gas connectors should be sealed with
PVC caps to prevent pollution by moisture,
dust, etc.
Fig. 4-2: Labelling of gas connectors (example)
Pressure control valve
Flow sensor
Exhaust
Exhaust
Analyzer
Sample gas pump
Filter
Fig. 4-3: Installation in bypass mode
The analyzer should be mounted close to the
sample gas source to minimize transport time.
A sample gas pump can be used to reduce the
reaction time; this requires that the analyzer
be operated in bypass mode or tted with
a pressure control valve to protect against
excessive gas ow and pressure (Fig. 4-3).
4.4 Installation - Gas Connections