Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG4-26
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
All signal cables are connected to screw-type
terminals located inside the housing. Access
to the internal components is gained by relea-
Preparation of signal cables
Recommended wire gauge: 0.14…1.5 mm
2
(AWG 26…AWG 16),
end sleeves not required
Skinning length: 5 mm (0.2")
Thread: M2
Min. tightening torque: 0.25 Nm (2.21 in.lb)
4.6.3 Installation - X-STREAM X2XF Field Housings
All cables must be fed through cable glands
and secured with a gland nut.
When correctly installed, the glands act as a
strain relief and guarantee EMC (electroma-
gnetic compatibility):
sing the two fasteners at the front door and
opening it to the left.
Gas inlets and outlets
Quantity: max. 8 (+ 1 purge gas inlet (optional))
Specication: 6/4 mm PVDF
optional 6/4 mm or
1
4", stainless steel, others on
request
Installing cable glands with shielded cables
3. Feed cable through
dome nut and clam-
ping insert
4. Fold braided shield
over clamping insert
5. Make sure that brai-
ded shield overlaps
the O-ring by
3
32
(2 mm)
1. Strip the cabel
2. Expose braided
shield
6. Push clamping insert
into body and tighten
dome nut
7. Assemble into housing
and you´re done!