Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 6-25
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6
Software Menus
6.2 Menu System - Valve Assignment Menu
6.2.3.2.2 Valve Assignment Menu
Setup..
In/Output..
Valve assignment..
Multi-channel unit:
Select the component to be set in the SELECT
COMPONENT menu.
If the system is set up accordingly, the access
code for level 3 must be entered to gain ac-
cess to this menu.
The internal and/or external valves are assi-
gned to zero or span gas in this menu (
"7.3 Calibration Procedures" at page 7-5).
Available options:
V1 ... V8
Valve assignment
Zerogas: V1
Spangas: V2
Multi-channel unit:
Press the key to enter the SELECT COM-
PONENT menu to change the settings for a
different channel.
Note for multi-channel units:
Valves can be freely assigned to channels.
This includes among others the following
variations:
• identical combinations (of zero and span
gas valves) for several channels
• Combinations in which one valve has the
same function for several channels
•
Combinations in which one valve has a dif-
ferent function depending on the channel;
e.g. a zero gas valve for channel 1 is also
the span gas valve for channel 2.
Depending on the gases use, the calibration
procedure can be optimised for time or efci-
ency using such combinations.
(
"7.3 Calibration Procedures" at page
7-5).