Emerson Process Management GmbH & Co. OHG 7-25
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
7
Maintenance & Procedures
7.4.3 Advanced Calibration
span all
channels with
the current
span gas valve
assigned
select
channel 1
increase
channel no.
by 1
select all
channels with
the same
span gas valve
assigned
is current
channel last
channel?
is this
channel
already
spanned?
start
span all
procedure
END
Yes
Yes
No
No
Fig. 7-5: Span All Calibration Procedure Flow Diagram