Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 5-35
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
5
Startup
5.7 Checking the Settings
The unit is now saving a copy of the current
conguration data (the so-called CfgData
dataset) in a special area of memory labelled
UserData. This dataset can be used to reset
the unit later if, for example, later incorrect
settings render the unit unusable.
If, during the analyzer startup up, the Cfg-
Data checksum is found to be incorrect, the
UserData dataset is loaded, to ensure the
instrument remains usable.
Further changes to the conguration will only
be stored in the CfgData dataset until manu-
ally saved to UserData.
Upon completion of the saving process a
conrmation message will be displayed.
Note!
For more detailed descriptions of all the ope-
rations in this menu:
7.6, page 7-51.
You have now completed checking the
analyzer setup: Press the MEASURE key
to return to the measurement display.
CfgData>UserData
Are you sure?
No!
Yes!
(i)
-COMMAND EXECUTED-
Copying data
- PLEASE WAIT -
Procedure X:E000
A new window comes up to conrm the action:
Select the line Yes! and press the ENTER
key: Another windows shows the current
status.