7.5.5 Replacing the Trace Moisture Sensor
Locate the sensor unit within your ana-
lyzer.
Open the ttings, connecting the unit to
the piping.
Only loosen the small nuts (do not remove
them), xing the sensor unit to the analy-
zer plate and carefully take the unit out of
the analyzer.
Take off the signal plug at the sensor's end
(it is xed by a screw!).
Place a wrench (size: 27 mm) at the hexa-
gon and screw out the sensor by turning
it counterclockwise (ccw).
Do not use the cylindric sen-
sor body to losen the sensor!
Place a second wrench with
size 30 mm from the top to
counterhold the block.
Take the new sensor and carefully place
the HDPE protected end of the sensor into
the sensor block.
Place the one wrench at the hexagon, the
other at the block to counterhold and x
the sensor by turning it clockwise (cw).
To ensure proper measure-
ments, apply a torque of
min. 30.5 Nm (269 in.lb).
Do not use the cylindric sensor
body to install the sensor!
Install the connector to the sensor and x
it with the screw.
Place the unit into the analyzer and x it
with the small nuts.
Re-install the piping ttings.
Make sure that all the plugs associated
with the sensor are properly connected
the same way as before.
30 mm
Connectors: 1 sensor signal plug
2 power to sensor
3 signal to analyzer
For details see appendix!
1
2
3
Consider all applicable safety
instructions, especially those at
the beginning of this section 7.5
Sensor block
Fig. 7-16: Trace Moisture Sensor Assembly Separated
7.5 Replacing Worn Out Sensors