Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG4-22
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
Pin Signal
P3.1 Input 1
P3.2 Input 2
P3.3 Input 3
P3.4 Input 4
P3.5 Input 5
P3.6 Input 6
P3.7 Input 7
P3.8 GND for inputs 1-7
P3.9 unused
P3.10 Output 5, NC
P3.11 Output 5, NO
P3.12 Output 5, COM
P4.1 Output 6, NC
P4.2 Output 6, NO
P4.3 Output 6, COM
P4.4 Output 7, NC
P4.5 Output 7, NO
P4.6 Output 7, COM
P4.7 Output 8, NC
P4.8 Output 8, NO
P4.9 Output 8, COM
P4.10 Output 9, NC
P4.11 Output 9, NO
P4.12 Output 9, COM
P2.1 Output 10, NC
P2.2 Output 10, NO
P2.3 Output 10, COM
P2.4 Output 11, NC
P2.5 Output 11, NO
P2.6 Output 11, COM
P2.7 Output 12, NC
P2.8 Output 12, NO
P2.9 Output 12, COM
P2.10 Output 13, NC
P2.11 Output 13, NO
P2.12 Output 13, COM
Digital inputs
Digital outputs
4.6.2 Installation - X-STREAM X2GP
Fig. 4-17: Conguration of XSTD terminal adapter
An XSTD adapter can optionally be used to
connect digital I/O cables to screw-type termi-
nals instead of Submin-D plugs and sockets:
it is plugged onto the X4.1 and X4.2 (if tted)
Submin-D connectors on the unit.
1 Connector for socket X4.1 / X4.2 (on reverse side)
2 Screw-type terminals
Note!
Consider the installation notes in section 4.5.
1 2
Recommended wire gauge:
0.14…1.5 mm
2
,
end sleeves not required
Skinning length:
5 mm
Thread:
M2
Min. tightening torque:
0.25 Nm
Note!
The conguration illustrated here is that of
the rst adapter (on socket X4.1).
Inputs 8-14 and outputs 14-23 are on the se-
cond adapter (on socket X4.2), if installed.