Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 9-33
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
9
Modbus Commands
Address Tag Name
Data
Type
Client
Access
Description
Daniel Modicon
3281…
3296
3281…
3296
Info.ManufacturingInfo String RO Infos stored for manufacturing purposes
3297 3297 Setup.Alarms.FlowAlarmTyp1 Word R/W alarm type ow limit comp1: 0=offFS 1=lowFS
3298 3298 Setup.Alarms.FlowAlarmTyp2 Word R/W alarm type ow limit comp2: 0=offFS 1=lowFS
3299 3299 Setup.Alarms.FlowAlarmTyp3 Word R/W alarm type ow limit comp3: 0=offFS 1=lowFS
3300 3300 Setup.Alarms.FlowAlarmTyp4 Word R/W alarm type ow limit comp4: 0=offFS 1=lowFS
3301 3301
Service.Measurement.DSPreset-
Count
Word R/W counter of DSP resets
3305 3305
Service.AnalogOut.AOutAdjust-
Start1
Short R/W offset value for ne adjustment of analog output1
3306 3306
Service.AnalogOut.AOutAdjust-
Start2
Short R/W offset value for ne adjustment of analog output2
3307 3307
Service.AnalogOut.AOutAdjust-
Start3
Short R/W offset value for ne adjustment of analog output3
3308 3308
Service.AnalogOut.AOutAdjust-
Start4
Short R/W offset value for ne adjustment of analog output4
3309 3309
Service.AnalogOut.AOutAdju-
stEnd1
Short R/W gain value for ne adjustment of analog output1
3310 3310
Service.AnalogOut.AOutAdju-
stEnd2
Short R/W gain value for ne adjustment of analog output2
3311 3311
Service.AnalogOut.AOutAdju-
stEnd3
Short R/W gain value for ne adjustment of analog output3
3312 3312
Service.AnalogOut.AOutAdju-
stEnd4
Short R/W gain value for ne adjustment of analog output4
3313 3313
Setup.Display.MeasureLines.
MeasLine5
Word R/W
signal assigned to line1 of 2nd measurement
display(0..n)
3314 3314
Setup.Display.MeasureLines.
MeasLine6
Word R/W
signal assigned to line2 of 2nd measurement
display(0..n)
3315 3315
Setup.Display.MeasureLines.
MeasLine7
Word
R/W
signal assigned to line3 of 2nd measurement
display(0..n)
3316 3316
Setup.Display.MeasureLines.
MeasLine8
Word R/W
signal assigned to line4 of 2nd measurement
display(0..n)
4001 4001 Status.NamurState Word RO
device‘s NAMUR state bit-eld (b0:F, b1:M, b2:O,
b3:C)
4002 4002 Status.DIO.PSAOutputState Word RO DOut state of a PSA Card
4003 4003 Status.Calibration.CalValveState Word RO
current state of the valves: bit0=sample, b1=V4,
b2=V1, b3=V2
4004 4004 Status.Calibration.CalibrationState Word RO
0=None 1=valve 2=purge 3=Zstart 4=Sstart
5=Zwait 6=Swait 7=cance
4005 4005 Status.Calibration.CalibrationCount Word RO
second decrementer for calibration and/or purging
procedures
9.4 List of Parameters and Registers - Sorted by Registers