Troubleshooting
8.3 Solving Problems Not Indicated by Status Messages
Situation Description Actions
Digital outputs not working
properly (cont.)
Outputs on extension board(s)
(XDIO) affected
XDIO: If LED "TIMEOUT" glows
- c
heck jumpers on XDIO.
XDIO #1: jumper on ADR2
XDIO #2: jumpers on ADR2 &
ADR0
XDIO: If LED "TIMEOUT" glows
- check connection to P33
XDIO: If LED "NO SPI" glows
- check internal SPI communi-
cation cable (10 pole cable)
Digital inputs not working
properly
External failure
Check external circuitry for
failures
Conguration failure
Check digital inputs menu
settings
Outputs on extension board(s)
(XDIO) affected
XDIO: If LED "TIMEOUT" glows
- c
heck jumpers on XDIO.
XDIO #1: jumper on ADR2
XDIO #2: jumpers on ADR2 &
ADR0
XDIO: If LED "TIMEOUT" glows
- check connection to P33
XDIO: If LED "NO SPI" glows
- check internal SPI communi-
cation cable (10 pole cable)
Internal valves not working
properly
Connection failure
Check electrical connection of
valves
XPSA: If red LED "TIMEOUT"
glows - check connection to
P33
XPSA: LED "TIMEOUT" dark
- check power connection to
XPSA (2-pole cable br/wht)
External valves not working
properly
Valves connected to digital
outputs
See "Digital outputs not wor-
king properly"
Valves not connected to digital
outputs
Check external valve control-
ler