Troubleshooting
8.3 Solving Problems Not Indicated by Status Messages
Situation Description Actions
Digital outputs not working
properly (cont.)
Outputs on extension board(s)
(XDIO) affected
XDIO: If LED "TIMEOUT" glows
- c
heck jumpers on XDIO.
XDIO #1: jumper on ADR2
XDIO #2: jumpers on ADR2 &
ADR0
XDIO: If LED "TIMEOUT" glows
- check connection to P33
XDIO: If LED "NO SPI" glows
- check internal SPI communi-
cation cable (10 pole cable)
Digital inputs not working
properly
External failure
Check external circuitry for
failures
Conguration failure
Check digital inputs menu
settings
Outputs on extension board(s)
(XDIO) affected
XDIO: If LED "TIMEOUT" glows
- c
heck jumpers on XDIO.
XDIO #1: jumper on ADR2
XDIO #2: jumpers on ADR2 &
ADR0
XDIO: If LED "TIMEOUT" glows
- check connection to P33
XDIO: If LED "NO SPI" glows
- check internal SPI communi-
cation cable (10 pole cable)
Internal valves not working
properly
Connection failure
Check electrical connection of
valves
XPSA: If red LED "TIMEOUT"
glows - check connection to
P33
XPSA: LED "TIMEOUT" dark
- check power connection to
XPSA (2-pole cable br/wht)
External valves not working
properly
Valves connected to digital
outputs
See "Digital outputs not wor-
king properly"
Valves not connected to digital
outputs
Check external valve control-
ler