Connecting the power cord
The power cord is connected to screw-type
terminals located inside the housing.
Supported wire cross sections: 0.2…4 mm
2
(24…12 AWG),
no need to use wire end sleeves
Cable skinning length: 8 mm (0.315 inch)
Hole diameter: 1.2 mm (0.05 inch)
Screw thread: M 3
Tightening torque, min: 0.5 Nm (4.4 in.lb)
Fig. 4-25: Power supply connections
4.6.3 Installation - X-STREAM X2XF Field Housings
ELECTRICAL SHOCK HAZARD
Verify the power supply at installation site meets the specication given on
the analyzer´s nameplate label, before installing the instrument!
Verify power cables are disconnected and/or instrument is de-energized
prior to working at the terminals!
Verify the power cord is layed with a distance of at least 1 cm (0.5“) to any
signal cable to ensure proper insulation from signal circuits!
Feed the power cable through the cable
gland at the instrument´s right side and strip
the outer insulation. Strip the individual wires
and connect to the terminals (a label is loca-
ted next to the terminals on the mains lter
housing).
Finally, tighten the outer dome nut to secure
the power cable.
Live L
Earth PE
Neutral N
Power supply
cable gland