7.4.3 Advanced Calibration
Gasow V2
Procedure Purging
Time 10 s
The procedure starts showing the screen
to the left, indicating that valve V2 is open.
Currently the system is purged (prepurge) to
ensure it is properly lled with span gas when
span calculation starts. Here the remaining
purge time is 10 seconds (decreasing from the
value entered in the calibration setup screen,
page 7-14).
Gasow V2
Procedure Spanning
Time 38 s
When the prepurge time has elapsed, the
instrument starts to calculate the span point
(indicated by the term Spanning): Span gas
is still applied, the count down starts at 40 s.
Note!
Multi channel instruments spanning more than
1 channel at a time, show Spanning in the
procedure line while calibration is ongoing. To
see which channels are currently calibrated,
see the status line in the measuring screen.
Gasow Sample
CO2.1 0.000 ppm
Procedure Purging
Time 10 s
After this calibration has nished, the instru-
ment is checked for additional channels not
yet span calibrated and if there are any, the
procedure starts again with the rst step as
described at the top of this page, now with
another span gas valve.
After span calibrations have nished, the
instrument closes all span gas valves and
opens the sample gas valve. Now a postpur-
ge procedure starts to indicate that proper
sample gas measurement values require the
system to be lled with the related gas only.
Postpurge time is purge time plus 2x damping
time as entered in the calibration setup (
page 7-14) and measurement setup (
page 6-29).
The span calibration procedure has nished
when the last time interval shows remaining
0 seconds and the gas ow is sample.
Now press the HOME key to return to the
measuring screen.