Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 5-1
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
5
Startup
Chapter 5
Startup
Once the unit has been unpacked and in-
stalled, it is recommended to rst check the
settings and if necessary adjust them to the
user’s needs. e.g:
• What hardware is installed?
• Is the unit configured to your needs
(alarms, inputs, outputs, etc.)
In order for the information in this chapter
to be of any relevence, the unit must have
been installed according to the instructions
in chapter 4.
5.1 Introduction
The following pages describe how to navigate
through the menus and what is to be observed
when conguring the unit. For the rst startup
after installation, follow the step-by-step in-
structions for navigating the menus, allowing
you to familiarise yourself with the unit and its
software, and if necessary adjust the settings
to your own requirements.
OPERATION AT LOW TEMPERATURES
When operating an instrument at temperatures below 0 °C (32 °F), do
NOT apply gas nor operate the internal pump before the warmup time has
elapsed!
Violation may result in condensation inside the gas paths or damaged pump
diaphragm!