Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG9-22
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
9.4 List Of Parameters And Registers - Sorted By Daniel Registers
9.4 List of Parameters and Registers - Sorted by Registers
Address Tag Name
Data
Type
Client
Access
Description
Daniel Modicon
2001 2001 Control.Calibration.Zero_1 Boolean R/W Zero calibration comp1 (1=start)
2002 2002 Control.Calibration.Zero_2 Boolean R/W Zero calibration comp2 (1=start)
2003 2003 Control.Calibration.Zero_3 Boolean R/W Zero calibration comp3 (1=start)
2004 2004 Control.Calibration.Zero_4 Boolean R/W Zero calibration comp4 (1=start)
2005 2005 Control.Calibration.Zero_All Boolean R/W Zero calibration all (1=start)
2006 2006 Control.Calibration.Span_1 Boolean R/W Span calibration comp1 (1=start)
2007 2007 Control.Calibration.Span_2 Boolean R/W Span calibration comp2 (1=start)
2008 2008 Control.Calibration.Span_3 Boolean R/W Span calibration comp3 (1=start)
2009 2009 Control.Calibration.Span_4 Boolean R/W Span calibration comp4 (1=start)
2010 2010 Control.Calibration.Span_All Boolean R/W Span calibration all (1=start)
2011 2011 Control.Calibration.ZeroSpan_All Boolean R/W Zero + span calibration all (1=start)
2012 2012
Control.Calibration.Calibration_
Cancel
Boolean R/W Cancel any calibration (1=cancel)
2013 2013
Setup.Calibration.RestoreCalibZe-
ro1
Boolean R/W
restore zero calibration parameters from user
memory for comp1
2014 2014
Setup.Calibration.RestoreCalibZe-
ro2
Boolean R/W
restore zero calibration parameters from user
memory for comp2
2015 2015
Setup.Calibration.RestoreCalibZe-
ro3
Boolean R/W
restore zero calibration parameters from user
memory for comp3
2016 2016
Setup.Calibration.RestoreCalibZe-
ro4
Boolean R/W
restore zero calibration parameters from user
memory for comp4
2017 2017
Setup.Calibration.RestoreCalib-
Span1
Boolean R/W
restore span calibration parameters from user
memory for comp1
2018 2018
Setup.Calibration.RestoreCalib-
Span2
Boolean R/W
restore span calibration parameters from user
memory for comp2
2019 2019
Setup.Calibration.RestoreCalib-
Span3
Boolean R/W
restore span calibration parameters from user
memory for comp3
2020 2020
Setup.Calibration.RestoreCalib-
Span4
Boolean R/W
restore span calibration parameters from user
memory for comp4
2021 2021 Control.ApplyGas.SampleValve Boolean R/W 0=close all valves, 1=open sample valve
2025 2025 Control.ApplyGas.ZeroValve1 Boolean R/W 0=open sample valve, 1=open zero valve comp1
2026 2026 Control.ApplyGas.ZeroValve2 Boolean R/W 0=open sample valve, 1=open zero valve comp2
2027 2027 Control.ApplyGas.ZeroValve3 Boolean R/W 0=open sample valve, 1=open zero valve comp3
2028 2028 Control.ApplyGas.ZeroValve4 Boolean R/W 0=open sample valve, 1=open zero valve comp4
2029 2029 Control.ApplyGas.SpanValve1 Boolean R/W 0=open sample valve, 1=open span valve comp1
2030 2030 Control.ApplyGas.SpanValve2 Boolean R/W 0=open sample valve, 1=open span valve comp2
2031 2031 Control.ApplyGas.SpanValve3 Boolean R/W 0=open sample valve, 1=open span valve comp3
2032 2032 Control.ApplyGas.SpanValve4 Boolean R/W 0=open sample valve, 1=open span valve comp4