Modbus Commands
9.4 List of Parameters and Registers - Sorted by Registers
Address Tag Name
Data
Type
Client
Access
Description
Daniel Modicon
3087 3087
Setup.In/Outputs.DigitalInOut.
DIO1In.Input7
Word R/W signal assigned to Input7 of DIO-board#1
3088 3088
Setup.In/Outputs.DigitalInOut.
DIO2In.Input1
Word R/W signal assigned to Input1 of DIO-board#2
3089 3089
Setup.In/Outputs.DigitalInOut.
DIO2In.Input2
Word R/W signal assigned to Input2 of DIO-board#2
3090 3090
Setup.In/Outputs.DigitalInOut.
DIO2In.Input3
Word R/W signal assigned to Input3 of DIO-board#2
3091 3091
Setup.In/Outputs.DigitalInOut.
DIO2In.Input4
Word R/W signal assigned to Input4 of DIO-board#2
3092 3092
Setup.In/Outputs.DigitalInOut.
DIO2In.Input5
Word R/W signal assigned to Input5 of DIO-board#2
3093 3093
Setup.In/Outputs.DigitalInOut.
DIO2In.Input6
Word R/W signal assigned to Input6 of DIO-board#2
3094 3094
Setup.In/Outputs.DigitalInOut.
DIO2In.Input7
Word R/W signal assigned to Input7 of DIO-board#2
3095 3095
Setup.Calibration.Auto.AutoZero-
TimeInterval
Word R/W
time interval in hours for automatic zero calibra-
tions
3096 3096
Setup.Calibration.Auto.AutoZero-
StartHours
Word R/W
hour part for next start of automatic zero calibra-
tion
3096 3096
Status.Calibration.AutoZeroStart-
Hours
Word RO
hour part for next start of automatic zero calibra-
tion
3097 3097
Setup.Calibration.Auto.AutoZero-
StartMinutes
Word R/W
minute part for next start of automatic zero calibra-
tion
3097 3097
Status.Calibration.AutoZeroStart-
Minutes
Word RO
minute part for next start of automatic zero calibra-
tion
3098 3098
Setup.Calibration.Auto.AutoZero-
SpanTimeInterval
Word R/W
time interval in hours for automatic zero&span
calibrations
3099 3099
Setup.Calibration.Auto.AutoZSpan-
StartHours
Word R/W
hour part for next start of automatic zero&span
calibrations
3099
3099
Status.Calibration.AutoZSpanStart-
Hours
Word RO
hour part for next start of automatic zero&span
calibrations
3100 3100
Setup.Calibration.Auto.AutoZSpan-
StartMinutes
Word R/W
minute part for next start of automatic zero&span
calibrations
3100 3100
Status.Calibration.AutoZSpanStart-
Minutes
Word RO
minute part for next start of automatic zero&span
calibrations
3101 3101 Setup.Alarms.Limit1AlarmTyp1 Word R/W
alarm type limit1 comp1: 0=off 1=lo 2=hi 3=offFS
4=loFS 5=hiFS
3102 3102 Setup.Alarms.Limit1AlarmTyp2 Word R/W
alarm type limit1 comp2: 0=off 1=lo 2=hi 3=offFS
4=loFS 5=hiFS
3103 3103 Setup.Alarms.Limit1AlarmTyp3 Word R/W
alarm type limit1 comp3: 0=off 1=lo 2=hi 3=offFS
4=loFS 5=hiFS