Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 4-5
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
4
Installation
4.4 Installation - Gas Connections
TOXIC GAS HAZARDS
Take care that all external gas pipes are connected in the described way
and that they are gastight to avoid leakages!
Faulty connected gas pipes lead to explosion hazard or even to mortal
danger!
Don‘t take a breath of the emissions! Emissions may contain hydrocarbons
or other toxic components (e.g. carbon monoxide)! Carbon monoxide may
cause headache, sickness, unconsciousness and death.
Do not confuse gas inlets and outlets. All gases supplied must be prepared
beforehand. When supplying aggressive gases, ensure that the gas lines
are not damaged.
Max. admissable pressure: 150 kPa / 7 psig; atmospheric with internal pump!
Exhaust lines must be installed to incline downwards and be unpressurized
and protected against frost, and conform to legal requirements.
4.4 Gas Connections
TRACE OXYGEN MEASUREMENTS
The sensor for trace oxygen measurements is a consumable.
Remaining lifetime counts down when the sensor is in contact with oxygen.
For above reasons, the analyzer is shipped with the sensor as extra item in a
sealed bag! The sensor must be installed before analyzer startup, according
the instructions shipped with the sensor!
Do not use plastic tubing for trace oxygen measurements as it can permeate
oxygen from the ambient air and cause higher than expected oxygen
readings.