Emerson Process Management GmbH & Co. OHG7-52
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
• Disconnect the analyzer from power
• Re-install the sensor unit into the analyzer
• Close the housing. Take care to use all
screws, especially if the instrument is to
be used in hazardous areas!
or
close the rear panel cover and secure it
with the screw nut.
7.5.3.4 Finalizing the Sensor Replacement
In a next step for proper measuring results,
perform a zero and a span calibration at least
for the channel with the replaced sensor.
To ensure proper disposal, send back the old
sensor to the EMERSON Process Manage-
ment factory (or to your local sales ofce) or
to an industrial waste management contractor
for waste disposal.
EXPLOSION HAZARD
Special conditions and instructions for start-up after maintenance apply
to instruments to be operated in hazardous areas!
Not observing these conditions and instructions may cause explosions!
See the associated manuals, provided with instruments for use in hazardous
areas, for more information.
7.5 Replacing Worn Out Sensors