7.4.3 Advanced Calibration
The analyzer immediately begins span
calibration(s). Watch the screen for status
information (in the following explained by
means of exemplary gures).
Gasow V4
Procedure Purging
Time 10 s
The procedure starts showing the screen
to the left, indicating that valve V4 is open.
Currently the system is purged (prepurge) to
ensure it is properly lled with span gas when
span calculation starts. Here the remaining
purge time is 10 seconds (decreasing from the
value entered in the calibration setup screen,
page 7-14).
Gasow V4
Procedure Spanning
Time 38 s
When the prepurge time has elapsed, the
instrument starts to calculate the span point
(indicated by the term Spanning): Span gas
is still applied, the count down starts at 40 s.
Note!
Multi channel instruments spanning more than
1 channel at a time, show Spanning in the
"procedure" line while calibration is ongoing.
To see which channels are currently cali-
brated, see the status line in the measuring
screen.
Gasow Sample
CO2.1 0.000 ppm
Procedure Purging
Time 10 s
After span calibration has nished, the instru-
ment closes the span gas valve and opens the
sample gas valve. Now a postpurge procedu-
re starts to indicate that proper sample gas
measurement values require the system to be
lled with the related gas only. Postpurge time
is purge time plus 2x damping time as entered
in the calibration setup ( page 7-14) and
measurement setup ( page 6-29).
The span calibration procedure has nished
when the last time interval shows remaining
0 seconds and the gas ow is sample.
Now press
either the LEFT key to return to the advanced
calibration menu to select another calibration
procedure
or the HOME key to return to the measuring
screen.