Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 6-55
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6
Software Menus
Gasow V2
Procedure Purging
Time 10 s
AutoCal in..
6.2 Menu System - Status Menu
6.2.4.5 Calibration Status
Status..
Calibration..
If the unit is currently being calibrated, this
menu provides information on the current sta-
tus. Unlike the CONTROL - ZERO/SPAN CA-
LIBRATION menus, for which a channel must
be selected for multi-channel units, the data
displayed here are not channel dependent;
i.e., the general status is displayed here.
Shows the current calibration status
Possible values:
None: no calibration in progress
Zeroing: zeroing a channel
Spanning: spanning a channel
Purging: purging gas lines
The third line indicates the time remaining for
the currently active procedure.
This submenu displays information about the
next autocalibration (
6.2.4.5.1, page
6-57)
Page 1
Gasow:
Possible values: Sample, V1...V8, None
These values represent the internal and/
or external valves used for autocalibration.
All valves except the sample valve are user
congurable: As each can be assigned either
zero or span gas, not the type of gas is listed
here but only the valve designator.