7.4.1 Preparing Calibrations
7.4.1.1.1 Internal Valve Assignment
If your analyzer provides internal valves, at
rst
open the SETUP - IN/OUTPUTS - INTSHS
menu to assign the valves to the gas inlets:
Setup..
In/Outputs..
IntSHS..
Gas5: Off
Gas6: V2
Gas7: Off
Gas8: Off
Page 2
Gas1: Sample
Gas2: V1
Gas3: V3
Gas4: Off
Page 1
The IntSHS menu allows to congure the
optional internal valves for routing gas.
Each available analyzer gas inlet ("Gas 1 ...
Gas 8") with a valve connected is assigned
a virtual valve label (V1...V8, Sample). (If
the components have been installed in the
factory, the conguration is already setup).
Notes!
Depending on the analyzer model, 1 or 2
valve blocks with up to 4 or 8 valves can be
installed.
The number of available gas connections
depends on the analyzer model and varies
from 4 to 8.
If already factory setup, changing the con-
guration could result in inproper operation!