Gas connections
Depending on the conguration of the unit
(number of measurement channels and serial
or parallel connection), sample and calibra-
tion gases are fed into the unit via up to 8
tube ttings mounted on the rear panel. The
conguration of the ttings is indicated on an
adhesive label located near the tube ttings.
Any free tube ttings can be used for purging
the device with
inert gas to minimize interference from the
ambient atmosphere when measuring small
concentrations (e.g. of CO
2
)
or
•
air or inert gas when measuring corrosive
and/or ammable gases.
For further information, see page 1-5.
1.5 X-STREAM X2GK
Special components
For the analysis of aggressive and/or am-
mable gases, infallible components (WLD;
photometer under development) can be used.
The risk of an uncontrolled release of gases
due to a leak can be minimized by the use
of stainless steel piping and clamping rings.
Further details about the infallible contain-
ment are in the separate manual supplied with
units containing such parts.
Furthermore, intrinsically safe measuring
cells, which prevent the ignition of gas mix-
tures in the case of a failure, can also be
used.
Modul variation XCC
The ½19-in unit is also available as an ana-½19-in unit is also available as an ana-
lyzer module without the front control panel,
which can, for example, be integrated into a
measuring system in which control and ana-
lysis of data is performed via interface by an
external data acquisition system.
1
Technical Description