Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG4-28
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
4.6.3 Installation - X-STREAM X2XF Field Housings
Fig. 4-21: Terminal block X1 - Analog signals and relay outputs 1-4
Note!
Consider the installation notes in section 4.5
and the notes on installing cable glands on
page 4-24.
Analog signals
Relay outputs
Terminals for analog signals and relais outputs
1 - 4 are located on the outer left module
(terminal block X1; Fig. 4-21).
Analog signal output specication: 4 (0)–20 mA; load: R
B
≤ 500 Ω
Specication of relay outputs 1-4:
dry relay change-over ontacts,
can be used as normally open (NO) or
normally closed (NC)
Electrical specication: max. 30 VDC, 1 A, 30 W
Pin Signal
P2.1 Channel 1, (+) 4 (0)–20 mA
P2.2 Channel 1, GND
P2.3 Channel 2, (+) 4 (0)–20 mA
P2.4 Channel 2, GND
P2.5 Channel 3, (+) 4 (0)–20 mA
P2.6 Channel 3, GND
P2.7 Channel 4, (+) 4 (0)–20 mA
P2.8 Channel 4, GND
P2.9 not used
P2.10 not used
P2.11 not used
P2.12 not used
P3.1 not used
P3.2 not used
P3.3 Output 1 (Failure), NC
P3.4 Output 1 (Failure), NO
P3.5 Output 1 (Failure), COM
P3.6 Output 2 (Maintenance Request), NC
P3.7 Output 2 (Maintenance Request), NO
P3.8 Output 2 (Maintenance Request), COM
P3.9 Output 3 (Out of Spec), NC
P3.10 Output 3 (Out of Spec), NO
P3.11 Output 3 (Out of Spec), COM
P3.12 Output 4 (Function check), NC
P4.1 Output 4 (Function check), NO
P4.2 Output 4 (Function check), COM
P4.3 not used
P4.4
P4.5
P4.6
P4.7
P4.8
P4.9
P4.10
P4.11
P4.12
Relay Outputs
**)
Analog Outputs
Serial Interface
*)
**)
Conguration of relay output termi-
nals as per standard factory setting
(NAMUR status signals)