6.2 Menu System - Measurement Setup
6.2.3.3 Measurement Menu
If no pressure sensor is installed (INSTALLED
OPTIONS - PRESSURE.. set to Manual), the
current ambient pressure must be set here.
Possible values: 500 .. 2000 hPa
Otherwise this line is not editable and shows
the currently measured pressure.
Setup signal damping
6.2.3.3.1, page 6-29
Note 1
The unit for pressure values is taken from the
relevant entry in the display setup.
Note 2
As the pressure value is used for pressure
compensation, it should, when set to
Manu-
al, be regularly updated to ensure accurate
results.
Setup..
Measurement..
If the system is set up accordingly, the access
code for level 2 must be entered to gain ac-