Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG 6-43
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6
Software Menus
6.2.3.5 Installed Options Menu
6.2 Menu System - Installed Options Menu
Setup..
Installed Options..
If the system is set up accordingly, the access
code for level 3 must be entered to gain ac-
cess to this menu.
Valves: Internal
COM-Interf: Yes
Pump: Yes
FlowAlm: Yes
Page 1
Indicates whether valves are installed.
Available options:
Internal, External, Int(ernal)+Ext(ernal),
None.
Indicates whether a serial interface is installed
(Yes) or not (No)
Indicates whether an internal pump is installed
(Yes) or not (No).
Page 2
DigitalIO: 1
Pressure: Internal
Analog outputs 4
More..
Indicates how many analog outputs are in-
stalled. Accepted values: 1 ... 4
Indicates whether digital I/O cards are in-
stalled (1, 1+2) or not (None)
Determines whether a barometric pressure
sensor is installed.
Available options:
Manual: No sensor available. Ambient at-
mospheric pressure must be entered manu-
ally (
6.2.3.3 page 6-28)
Internal: Internal pressure sensor installed.
External: Pressure sent via network (e.g.
DeltaV)
More sensors
(
6.2.3.5.1 page 6-44).
Indicates whether ow switches are installed
(Yes) or not (No).