Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG4-36
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
X-STREAM X2GP with screw-type terminal
adapters
In order to avoid measured values being inu-
enced by external interference signals when
terminal adapters are in use, the signal cable
shieldings must be connected to the analy-
zer housing by means of shield connector
terminals:
Strip the signal cable to a length of 20 cm
(8"). Take care to not damage the braided
shield!
Pull up the contact part of the shield con-
nector terminal,
feed through the cable as illustrated in
g. 4-30,
release the contact part down onto the
braided shield.
This results in a secure contact with the cable
shielding and improves the unit’s interference
resistance.
The individual wires are then connected as
described in section 4.5.2.
The shield connector must be ordered to t
the cable diameter, and can be retrotted:
Ø 1.5 - 6.5 mm: part # ETC02019
Ø 5 - 11 mm: part # ETC02020
Ø 10 - 17 mm: part # ETC02021
Ø 16 - 24 mm: part # ETC02022
Fig. 4-29: Shield connector terminal with cable
4.7 Installation - Notes on Wiring