Software Menus
6.2 Menu System - Status Menu
6.2.4.1 Failures
Status..
1 Failures..
The number in the “Failures” line in the pre-
vious menu indicates how many errors are
currently active (here: 1). In the following
menus, Yes is displayed in the correspondung
number of lines.
For more details on troubleshooting,
Chapter 8 Troubleshooting.
Note!
If one or more failures are active, the left-hand
LED on the front panel ashes red. The corre-
sponding relay is also activated, if congured
for NAMUR error messages.
Multi-channel unit:
Select the component in the SELECT COM-
PONENT menu.
Multi-channel unit:
Press the key to enter the SELECT COM-
PONENT menu to select a different chan-
nel.
ROMmemory Yes
SensTimeout No
SensCmdFail No
ADC-Error No
Chopper No
Detector No
Source No
Page 2
(e.g.: a ROM memory error is indicated)