Message
Status level
Explanation
Description Actions
SCalRefused
Check request
Explanation:
This message does not
appear during normal
operation modes!
Wrong DSP (sensor) congu-
ration
Call for service
FlowLimAlm
Check request
Explanation:
Flow too low
Flow sensor detected ow to
be below set limit
Increase ow
Set new limit
TempRange
Off-spec
Explanation:
Only at instruments with
thermostate control
Temperature during operation
out of range
Check heater elements
TempSensor
Off-spec
Explanation:
Temperatur sensor
defective
Instruments µP detected tem-
perature sensor to be defec-
tive
Check temperature sensor
Alarm Level1
- -
Explanation:
Concentration alarm level
1 is activated (exceeded)
Alarm level 1 was exceeded
Adjust the gas concentration
to be within the set limits
Alarm Level2
- -
Explanation:
Concentration alarm level
2 is activated (exceeded)
Alarm level 2 was exceeded
Adjust the gas concentration
to be within the set limits
8.2.2 Channel Related Messages