Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG9-30
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
9.4 List of Parameters and Registers - Sorted by Registers
Address Tag Name
Data
Type
Client
Access
Description
Daniel Modicon
3169 3169
Setup.Display.MeasureLines.
MeasLine4
Word R/W
signal assigned to line4 of 1st measurement
display(0..n)
3170 3170
Setup.Display.MenuAccesss.Auto-
CodeMode
Word R/W
denes how codes autom. locked
(0=never,1=home,2=1minute)
3171…
3174
3171…
3174
Setup.Display.MenuAccesss.Basi-
cAccessCode
String R/W user code for getting access to basic areas
3175 3175
Setup.Display.MenuAccesss.Basi-
cAccMode
Word R/W
mode for access to basic areas
(0=allowed,1=code, 2=prohibited)
3176…
3179
3176…
3179
Setup.Display.MenuAccesss.Exper-
tAccessCode
String R/W user code for getting access to expert areas
3180 3180
Setup.Display.MenuAccesss.Exper-
tAccMode
Word R/W
mode for access to expert areas (0=allowed,1=co
de,2=prohibited)
3181…
3184
3181…
3184
Setup.Display.MenuAccesss.Spe-
cialAccessCode
String R/W user code for getting access to special areas
3185 3185
Setup.Display.MenuAccesss.Spe-
cialAccMode
Word R/W
mode for access special areas (0=allowed,1=code
,2=prohibited)
3186…
3189
3186…
3189
Service.General.EmersonAccCode String R/W code for getting LOI access to Emerson areas
3190 3190 Service.General.EmersonAccMode Word RO mode for getting LOI access to Emerson areas
3191 3191 Service.Measurement.Simulation Word R/W
biteld for enabling simulation of single compon-
ents
3192 3192
Service.Measurement.StartRaw-
Meas1
Word R/W start raw measurement of component1
3193 3193
Service.Measurement.StartRaw-
Meas2
Word R/W start raw measurement of component2
3194 3194
Service.Measurement.StartRaw-
Meas3
Word R/W start raw measurement of component3
3195 3195
Service.Measurement.StartRaw-
Meas4
Word R/W start raw measurement of component4
3196…
3200
3196…
3200
Info.SerialNumber String RO serial number of the device
3196…
3200
3196…
3200
Service.General.SerialNumber String R/W serial number of the device
3201…
3210
3201…
3210
Info.ChannelId1 String RO channel identication text of comp1
3201…
3210
3201…
3210
Service.General.ChannelId1 String R/W channel identication text of comp1
3211…
3220
3211…
3220
Info.ChannelId2 String RO channel identication text of comp2
3211…
3220
3211…
3220
Service.General.ChannelId2 String R/W channel identication text of comp2