Maintenance & Procedures
7.4.6 Resetting a Calibration
7.4.7 Verifying a Calibration
7.4.6 Resetting a Calibration
In case a wrong conguration was detected
after calibration was performed (e.g. wrong
gas connected) there is an option to restore
the last user saved calibration data:
Within the screen where to start the calibrati-
on (either for span or zero) open the second
page:
A new screen appears with the option RESET
CALIBRATION.. Pressing the ENTER key in
this line results in a prompt for conrmation.
Choosing YES! replaces the current calibrati-
on data with the last calibration data, restored
from UserData;
7.6.2 SAVE-LOAD, page
7-54.
CANCEL calibration!
START calibration!
ZeroGas 0.000 ppm
CO2.1 0.200 ppm
RESET calibration..
CO2.1 0.200 ppm
RESET calibration
Are you sure?
No!
Yes!
Starting from the measurement screen press
the DOWN key to open the MAIN MENU,
enter the CONTROL.. menu.
Enter the last line (APPLY GAS..)
For instruments without internal and/or
external valves simply apply either span or
zero calibration gas to the sample gas inlet.
If the calibration still is proper, the reading
on the measurement screen should show the
related value.
For instruments with internal and/or external
valves follow the procedure below:
Zero calibration..
Span calibration..
Adv. calibration..
Apply gas..
Control..
Multi-channel unit:
Select the component to be veried in the
SELECT COMPONENT menu.