8.4 Troubleshooting on Components
Fig. 8-3: X-STREAM X2 Field housings and X2FD - How to Open
Sash fasteners
Screws at the ange
8.4.1.4 How To Open X-STREAM X2FD
To open a X-STREAM X2FD loosen the 20
screws located at the instrument´s ange.
Then carefully ip down the front door to not
damage the instrument, hinges or equipment
installed below the analyzer.
8.4.1.3 How To Open X-STREAM Field-
housings
EXPLOSION HAZARD
X-STREAM X2FD as well as special variations of X-STREAM XLF and XXF
Fieldhousings are intended to be installed in hazardous areas.
Maintaining such instruments is permitted only considering special
conditions, givenin the associated separate manuals.
Do not open nor maintain instruments in hazardous areas without having
read and understood all related instruction manuals!
Depending on the individual analyzer con-
guration, either open the upper or lower
front door to the left, utilizing the two sash
fasteners.
GASKETS AT LOW TEMPERATURES
Consider that enclosure gaskets may be frozen if the instrument is
installed outdoors. Carefully open the enclosure at temperatures below
-10 °C to not damage the gaskets.
Damaged gaskets void the ingress protection, possibly causing property
damage, personal injury or death.