Emerson Process Management GmbH & Co. OHG6-8
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6.2.2.2.1 RESET Calibration Menu
6.2 Menu System - Control Menu
RESET
Are you sure?
No!
Yes!
This menu appears if the user selected the
menu line "Reset.." in either the ZERO CA-
LIBRATION or SPAN CALIBRATION menu.
To reset the current calibration data to the va-
lues saved in UserData (
6.2.3.8 SAVE-
LOAD menu on page 6-48), select "Yes!" and
press the 8 key.
"No!" returns to the previous menu without
applying changes to the calibration data.
Note!
Unless altered by the user, UserData settings
are the same as the factory settings.
Note 2!
Once the reset procedure is started, a mes-
sage is displayed indicating how to abort the
procedure.