Emerson Process Management HASX2E-IM-HS Carbon Monoxide Alarm User Manual


 
Emerson Process Management GmbH & Co. OHG6-34
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
6.2 Menu System - Analog Outputs Menu
6.2.3.4.1.2 Output1 (2 ... 4) Menu
Selects the parameter to be displayed in the
corresponding line in the measurement dis-
play. Depending on the number of measuring
channels and sensors installed in the unit, the
following values are available:
Gas components: Comp-1 ... Comp-4
Temperature: Temp-1 ... Temp-4,
Pressure: Press-1 ... Press-4,
Flow: Flow-1 ... Flow-4
Zoom: Zoom-C1 ... Zoom-C4
(nothing): None
A constant signal of either 0/4 mA or 20 mA
can be generated to check the output settings.
The options are labelled accordingly (0/4 mA
and 20 mA).
Setup..
In/Outputs..
Analog outputs..
Output1(2...4)..
Signal: Comp-1
LowScale: 10 ppm
HighScale: 1000 ppm
Zoom..
This line determines which concentration
value corresponds to the lower signal limit
(0 or 4 mA).
This line determines which concentration
value corresponds to the upper signal limit
(20 mA).
This submenu sets the zoom function for the
analog output;
6.2.3.4.1.2.1, page 6-36
Note!
The last line only appears when "signal" is
set to a zoom value (e.g. Zoom-C1).
Page 1